Specification : |
|
▪ Measuring principle |
: UV absorption (CVAAS) |
▪ Wavelength |
: 253,7 nm
|
▪ UV source |
: Electrodeless low-pressure mercury lamp (EDL)
|
▪ Stabilization method |
: Reference beam method
|
▪ Optical cell |
: Fused silica (Suprasil)
|
▪ Measuring ranges: |
: 0.1 - 100 µg/m³ |
|
: 0 - 1000 µg/m³
|
|
: 0 - 2000 µg/m³
|
▪ Sensitivity |
: 0.1 µg/m³ (0.01 ppb) |
▪ Response time |
: Approx. 1 second, real time measurement |
▪ Alarm |
: Exceedance of concentration, 3 levels programmable |
▪ Status alarms |
: measuring cell soiled |
|
: battery state
|
|
: UV lamp exhausted |
▪ Control pad |
: waterproof membrane keypad
|
▪ Measurement Display |
: 5.7 inch TFT graphical color display, with LED backlight |
▪ Signal outputs |
: USB and Bluetooth
|
▪ Software for data transfer |
: Mercury Instruments "Hg-Transfer" included on delivery, |
|
export of data: EXCEL® and ASCII format |
▪ Data storage |
: built-in Data logger, internal 4 GB SD-card |
|
automatic time and date stamp, GPS stamp if activated |
▪ GPS |
: built-in, receives US-GPS, Galileo, GLonASS and others |
|
for increase in positioning accuracy and improved reception |
▪ Internal power supply |
: rechargeable 12 V Eneloop batteries with min. Capacity |
|
of 7 hours and minimized self-discharge, quick swap battery
|
|
packs as option |
▪ External power supply |
: 110 - 230 VAC/50 - 60 Hz; with mains adapter external
|
|
12 V DC sources, (Cigarette lighter adapter optional)
|
▪ Dimensions |
: 29 x 17 x 14 cm (11.6" x 6.8" x 5.6") (W x H x D) |
▪ Weight |
: 2.6 kg (incl. battery pack)
|
▪ Complete unit |
: Certificate calibration from manufacturing |
|